基本信息
文件名称:JEOL 2010F说明书用户手册.pdf
文件大小:104.57 KB
总页数:6 页
更新时间:2025-05-16
总字数:约1.12万字
文档摘要

JEOL2010FMANUAL

Quickchecklist

1.Checktheacceleratingvoltage,mustbeat200kV(rightscreen),HT(μA)(leftpanel)at

0.96-0.97andEmission(leftpanel)at155-160.

2.Checkthevacuumsequence(leftdrawer),V17,V13,V5BandV2on.Vacuummeasured

onSIP(SIPunit,secondvacuummeterfromthetop)mustbebetterthan2x10-5Pa(blue

numbers,150l/s).Thevacuuminthegun(SIPunit,vacuummeteratthetop)mustbebetter

thanx10-6Pa.

3.Ifneeded,fillthereservoirwithLN2.Besurethattheviewingscreencoverisonbefore

fillingthetrap.

4.InsertspecimenholderintoTEM(Insertholderinairlock,setair/pumpswitchtopump

(yellowlightturnsON),whengreenlightturnsON,rotatefullyclockwiseintwostagesand

pushincompletely)

5.Checkthevacuumsequenceagain,(leftdrawer),V17,V13,V5B,V8,V21andV2on.

VacuummeasuredonSIP(SIPunit,secondvacuummeterfromthetop)mustbebetterthan2

x10-5Pa(bluenumbers,150l/s).Thevacuuminthegun(SIPunit,vacuummeteratthetop)

mustbebetterthanx10-6Pa.

6.Ifvacuumisok,openthevalvebetweenthecolumnandthegunbyclickingbutton

VALVE(leftpanelcontrols).

7.Ifneeded,alignthemicroscope.Work!

1.Initialcheck

1.Checkthevacuumsequence(leftdrawer),V17,V13,V5BandV2on.Vacuum

measuredinSIP(SIPunit,secondvacuummeterfromthetop)mustbebetterthan2x

-5Pa(bluenumbers,150l/s).Thevacuuminthegun(SIPunit,vacuummeteratthe

10

top)mustbebetterthanx10-6Pa.

2.Checkthecoolingwaterflowmetersonthebackwallinthemaintenanceroom.

3.Ifneeded,filltheanti-contaminationdevice(ACD)(coldtrap)withLN2.Besurethat

theviewingscreencoverisonbeforefillingthetrap.

4.Caution:TheDewarshouldbere-filledevery4-6hours.

2.InsertingthespecimenholderintheTEMcolumn

Loadthesample

?UseGloves!Lo